R.L. Snyder and K.T. Paw U
Copyright Regents of the University of California
Created January 6, 2002
Copyright
2002 Regents of the University of
California
These
teaching notes discuss psychrometric relationships that are important
for
understanding humidity and its measurement.
The first section discusses concepts related to measuring
dry-bulb
temperature. Then concepts pertaining
to thermodynamic wet-bulb temperature are presented.
Finally, some ideas on measuring the wet bulb temperature are
discussed.
The
heat transfer to and from a dry-bulb thermometer depends on the
long-wave radiation
balance and sensible heat transfer.
Assuming an emissivity e = 1.0, the net long-wave
radiation can be expressed in terms of a resistance to radiation
transfer (rR) in s m-1.
(1)
Ts is the shield temperature
(K) and Tt is the
thermometer temperature (K). r is the air density (kg m-3) and Cp is the specific heat at
constant pressure (kJ kg-1 oC-1). The transfer of
sensible heat by convection to and from the dry-bulb thermometer inside
the
shield is given by:
(2)
H is the sensible heat to and
from the thermometer and T is the air
temperature within the shield. When in
equilibrium, Rn = H, so
(3)
(4)
The
goal is to have Tt be very
close to T. This is
accomplished by making:
(1)
rH <<< rR
by
using a very small thermometer or by ventilating more
(2)
Ts very close to T by painting
the shield white, by insulating between outer and inner surfaces, or by
increasing ventilation on both sides of the screen.
In
an adiabatic system, the sum of latent and sensible heat remains
constant. The initial, energy state is
specified by
the temperature (T), vapor pressure (e),
and total pressure (p).
If liquid water is present and e
is smaller than es(T), then
water will evaporate and both e and p
will increase. Assuming
no exchange of heat between the system and the outside environment, an
increase
of latent heat in the system, represented by an increase in e,
must be balanced by a decrease in
sensible heat, represented by a decrease in T. This process will continue until the air
becomes saturated at T' (the
thermodynamic wet-bulb temperature).
The corresponding saturation vapor pressure is es(T'). The initial water vapor
density, defined by T and e, is
kg
m-3
(5)
where,
Mv and Md are
the molecular weights
(kg mol-1) of dry air and water vapor, respectively, e is the vapor pressure, p is the
barometric pressure, and r is the dry air density.
When the vapor pressure rises from e to es(T'), the change in latent
heat per unit
volume is
kJ
m-3
(6)
where
l is the latent heat of
vaporization (kJ kg-1). The
corresponding sensible heat used for vaporizing the water is
kJ
m-3
(7)
Equating
the two expressions and solving for e,
we get
kPa
(8)
where
is the psychrometric
constant. Actually, g is not a constant but a function of
barometric
pressure (p), which depends on
elevation and passing weather systems and on the latent heat of
vaporization (l), which is a weak function of temperature. If p
= 101.3 kPa, then g = 0.066 at 0oC
and g = 0.067 at 20oC.
A good formula to estimate p is
kPa
(9)
To
estimate l, use
J
kg-1
(10)
Note that vapor pressure can also be determined from es(T) rather than es(T'). Using the slope of the saturation vapor pressure curve at the mean of the air and wet-bulb temperatures (D’), small changes in saturation vapor pressure are expressed as:
kPa
(11)
Substituting
into the psychrometric equation (Eq. 8), we get
kPa
(12)
Substituting
the slope of the saturation vapor pressure curve at air temperature (D) for D’, and solving for T'
provides an equation to estimate the wet-bulb temperature.
oC
(13)
The
error in using D rather than D’ is small when T T'
is small, but the error increases with greater wet-bulb depression.
The
measured wet-bulb temperature (Tw)
is an estimate of the thermodynamic wet-bulb temperature (T'). For a wet-bulb
thermometer with temperature (Tw)
when exposed to air at temperature (T)
and surrounded by a radiation shield at air temperature, the rate of
sensible
heat gain (H) and net radiation (Rn)
can be expressed as
kJ
m-2s-1
(14)
where rHR is the parallel resistance to convective and radiation heat transfer. The rate of latent heat loss from the wet-bulb is
kJ m-2s-1
(15)
Recall that the psychrometric constant is
so
and
(16)
By substitution, we get
(17)
In equilibrium, lE = Rn + H, so
(18)
then
(19)
and
(20)
Thus, the wet-bulb temperature equals the thermodynamic wet-bulb temperature only when rv = rHR. The actual equation to estimate e is
(21)
where
. Based on the
concepts of forced convection, the resistance to vapor transfer equals
93% of
the resistance to convective heat transfer (rv
= 0.93 rH) for an
aspirated wet-bulb thermometer at 20oC.
When g
= g*, then
the resistance to vapor transfer
equals the parallel resistance to convective and radiative heat
transfer. Therefore,
(22)
Rearranging terms, we get
(23)
and
(24)
For example, if rR=210 s m-1, then rH=17 s m-1. Then the measured wet-bulb temperature is bigger than the thermodynamic wet-bulb temperature if rH>17 s m-1. The measured wet-bulb is smaller when rH <17 s m-1. Therefore, if a thermometer with rR=210 s m-1 is ventilated so that rH»17 s m-1, the measured will approximately equal the thermodynamic wet-bulb temperature. The radiation resistance is mainly affected by the thermometer size and shielding. Resistance to sensible heat transfer is mainly affected by ventilation.